Necessity for surgical revision of defibrillator leads implanted long-term : causes and management

Eckstein, J. and Koller, M. T. and Zabel, M. and Kalusche, D. and Schaer, B. A. and Osswald, S. and Sticherling, C.. (2008) Necessity for surgical revision of defibrillator leads implanted long-term : causes and management. Circulation, Vol. 117, H. 21. pp. 2727-2733.

Full text not available from this repository.

Official URL: http://edoc.unibas.ch/dok/A6006061

Downloads: Statistics Overview

Abstract

BACKGROUND: Defibrillator lead malfunction is a potential long-term complication in patients with an implantable cardioverter-defibrillator (ICD). The aim of this study was to determine the incidence and causes of lead malfunction necessitating surgical revision and to evaluate 2 approaches to treat lead malfunction. METHODS AND RESULTS: We included 1317 consecutive patients with an ICD implanted at 3 European centers between 1993 and 2004. The types and causes of lead malfunction were recorded. If the integrity of the high-voltage part of the lead could be ascertained, an additional pace/sense lead was implanted. Otherwise, the patients received a new ICD lead. Of the 1317 patients, 38 experienced lead malfunction requiring surgical revision and 315 died during a median follow-up of 6.4 years. At 5 years, the cumulative incidence was 2.5% (95% confidence interval, 1.5 to 3.6). Lead malfunction resulted in inappropriate ICD therapies in 76% of the cases. Implantation of a pace/sense lead was feasible in 63%. Both lead revision strategies were similar with regard to lead malfunction recurrence (P=0.8). However, the cumulative incidence of recurrence was high (20% at 5 years; 95% confidence interval, 1.7 to 37.7). CONCLUSIONS: ICD lead malfunction necessitating surgical revision becomes a clinically relevant problem in 2.5% of ICD recipients within 5 years. In selected cases, simple implantation of an additional pace/sense lead is feasible. Regardless of the chosen approach, the incidence of recurrent ICD lead-related problems after lead revision is 8-fold higher in this population.
Faculties and Departments:03 Faculty of Medicine > Departement Klinische Forschung > Clinical Epidemiology and Biostatistics CEB > Klinische Epidemiologie (Bucher H)
03 Faculty of Medicine > Bereich Medizinische Fächer (Klinik) > Kardiologie
03 Faculty of Medicine > Departement Klinische Forschung > Bereich Medizinische Fächer (Klinik) > Kardiologie
03 Faculty of Medicine > Bereich Medizinische Fächer (Klinik) > Kardiologie > Kardiologie Elektrophysiologie (Osswald)
03 Faculty of Medicine > Departement Klinische Forschung > Bereich Medizinische Fächer (Klinik) > Kardiologie > Kardiologie Elektrophysiologie (Osswald)
UniBasel Contributors:Osswald, Stefan and Sticherling, Christian and Koller, Michael
Item Type:Article, refereed
Article Subtype:Research Article
Publisher:Lippincott Williams & Wilkins
ISSN:0009-7322
Note:Publication type according to Uni Basel Research Database: Journal article
Related URLs:
Identification Number:
Last Modified:08 Nov 2012 16:23
Deposited On:08 Nov 2012 16:18

Repository Staff Only: item control page