edoc-vmtest

Undetectable phenytoin serum levels by an automated particle-enhanced turbidimetric inhibition immunoassay in a patient with monoclonal IgM lambda

Brauchli, Yolanda B. and Scholer, André and Schwietert, Martin and Krähenbühl, Stephan. (2008) Undetectable phenytoin serum levels by an automated particle-enhanced turbidimetric inhibition immunoassay in a patient with monoclonal IgM lambda. Clinica chimica acta, Vol. 389, H. 1-2. pp. 174-176.

Full text not available from this repository.

Official URL: http://edoc.unibas.ch/dok/A6005481

Downloads: Statistics Overview

Abstract

Phenytoin is a drug used for the treatment of different types of seizures. Its variable pharmacokinetics, mainly a consequence of variable bioavailability, saturable protein binding and saturable hepatic metabolism, predisposes the drug to therapeutic drug monitoring. Several methods to analyze the drug in serum exist with immunoassays being the method of choice for routine measurements. Immunoassays are specific and sensitive, but cross-reactivity, possibly leading to erroneous serum levels, is a concern. We report a patient with falsely undetectable phenytoin serum levels.
Faculties and Departments:03 Faculty of Medicine > Bereich Medizinische Fächer (Klinik) > Klinische Pharmakologie > Klinische Pharmakologie (Krähenbühl)
03 Faculty of Medicine > Departement Klinische Forschung > Bereich Medizinische Fächer (Klinik) > Klinische Pharmakologie > Klinische Pharmakologie (Krähenbühl)
UniBasel Contributors:Krähenbühl, Stephan
Item Type:Article, refereed
Article Subtype:Research Article
Publisher:Elsevier
ISSN:0009-8981
Note:Publication type according to Uni Basel Research Database: Journal article
Last Modified:07 Dec 2012 13:03
Deposited On:07 Dec 2012 13:01

Repository Staff Only: item control page