Oosterkamp, T. H. and Poggio, M. and Degen, C. L. and Mamin, H. J. and Rugar, D.. (2010) Frequency domain multiplexing of force signals with application to magnetic resonance force microscopy. Applied Physics Letters, 96 (8). 083107.
|
PDF
- Published Version
339Kb |
Official URL: http://edoc.unibas.ch/48612/
Downloads: Statistics Overview
Abstract
Frequency domain multiplexing, using an actively damped micromechanical cantilever, is used to detect multiple force signals simultaneously. The measurement principle is applied to magnetic resonance force microscopy to allow concurrent measurement of nuclear spin signals originating from distinct regions of the sample, or from multiple spin species.
Faculties and Departments: | 05 Faculty of Science > Departement Physik > Physik > Nanotechnologie Argovia (Poggio) |
---|---|
UniBasel Contributors: | Poggio, Martino |
Item Type: | Article, refereed |
Article Subtype: | Research Article |
Publisher: | AIP Publishing |
ISSN: | 0003-6951 |
e-ISSN: | 1077-3118 |
Note: | Publication type according to Uni Basel Research Database: Journal article -- The final publication is available at AIP Publishing, see DOI link |
Language: | English |
Identification Number: |
|
edoc DOI: | |
Last Modified: | 30 Jan 2017 15:54 |
Deposited On: | 30 Jan 2017 15:54 |
Repository Staff Only: item control page